The PMX3400 Three Mode Continuous & Intermittent model series provides reliable and accurate cycling of the regulator On and Off during the intermittent suctioning. The Three Mode Continuous & Intermittent series has a peak vacuum limit feature to prevent adjusting the vacuum too high. This series is ideal for sensitive applications where limiting Max vacuum is critically important. The intermittent timing cycle is factory set at 16 seconds on and 8 seconds off.
The unique modular design makes repair a breeze for any biomed or central supply department. The PMX3400 Three Mode Continuous & Intermittent is made up of eight (8) easy-to-replace modular components.
Unlike the normal plastic inlet and outlet connections on most units, the cast metal backplate connections are extremely durable, thereby saving unnecessary repair costs.
The case is made of strong plastic. It features a smooth surface with rounded corners making cleaning easy. The case also houses the gauge for extra protection.
Large gauge size makes it easy to see suction pressure. The PMX Series gauge is 50% larger than the PM Series gauge. The large control knob with its three (3) easy-to-read settings make vacuum adjustments simple, even in emergency situations.
The Pediatric vacuum regulator is restricted to 170 mmHg (±10 mmHg) to prevent high vacuum situations and the gauge face is imprinted with the Precision Medical’s “Stork Symbol” to identify the unit for pediatric applications.
Regulator Mode: Continuous-Intermittent 3 Mode (0-150 mmHg)
Custom colors are available. Please call customer service at 1-800-272-7285 for details.
Caution! U.S. Federal law restricts this device to sale by or on the order of a physician.
Depth: 4.5 in (11.4 cm)
Width: 3.8 in (9.5 cm)
Height: 5.7 in (14.4 cm)
Weight: 1.1 lbs (.50 kg)
PMX3400: 0 – 150 mmHg – Full Vacuum*PMX3400E: 0 – 150 mmHg (0 – 26 kPa)
*Counterclockwise direction
Analog: ± 5% of MAXDigital Gauge: ± 1% of Full Scale
Analog Gauge: ± 5% of MAX within ref. Indicator
1/8″ NPT Female
REG. (Regulated): provides an adjustable, continuous vacuum levelOFF: No Vacuum
INT. (Intermittent): provides an adjustable vacuum level that cycles between ON and OFF
REG: 50 l/min
INT: 10 l/min
*Maximum flow is obtained with a vacuum source of 21″ Hg (71.1 kPa)
REG. Mode @ Full Vac – 396 mmHg (53 kPa)
Factory set at sixteen (16) seconds ON and eight (8) seconds OFF (Reference only)
0°F to 122°F (-18°C to 50°C)
55°F to 85°F (13°C to 29°C)
Temperature Range:
-4°F to 140°F (-20°C to 60°C)Humidity:
Max 95% Non-condensing
Digital Vacuum Gauge Models ONLY
3 Volt Lithium, ½ AA
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Precision Medical, Inc. warrants that the Large Gauge Medical Vacuum Regulator (the Product) will be free of defects in workmanship and/or material for the following period:
Ten (10) years from date of shipment.
Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representative of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of nonconformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.