Helium Oxygen Blender

$1,790.54

SKU: Helium Oxygen Blender

Image may not reflect product configuration.

The HeliO2 Helium-Oxygen Blender is the only freestanding pneumatic HeliO2 blending device on the market today. The ability to turn the bleed control collar ON or OFF provides ± 3% FiO2 accuracy as well as quieter operation and maximal heliox tank durations.

The tough housing makes the unit durable and the selection control knob is recessed to prevent inadvertent changes in FiO2. The modular design can reduce maintenance time and costs. The unit can be easily maintained by all biomed departments.

The unit is available in four models, High-Flow 80/20 or 70/30 and Low-Flow 80/20 or 70/30. When paired with the optional cylinder clamp and pole, clinician setup time is greatly reduced and tank changes are easy. Multiple accessories can conform to clinician and user needs.

Features:

  • The world’s only freestanding pneumatic helium oxygen blender
  • Modular and customizable
  • Multiple accessories available for suiting various clinical application
Primary Outlet Flow Range
High Flow
PM5580: 80/20 Heliox
PM5570: 70/30 Heliox
15 – 120 l/min
supply pressures at 50 psi (3.4 bar) with BLEED Closed

Low Flow
PM5480: 80/20 Heliox
PM5470: 70/30 Heliox
3 – 30 l/min
supply pressures at 50 psi (3.4 bar) with BLEED Closed

Auxiliary Outlet Flow Range
High Flow
PM5580: 80/20 Heliox
PM5570: 70/30 Heliox
2 – 100 l/min
supply pressures at 50 psi (3.4 bar) with BLEED Closed

Low Flow
PM5480: 80/20 Heliox
PM5470: 70/30 Heliox
0 – 30 l/min
supply pressures at 50 psi (3.4 bar) with BLEED Closed

Bleed Flow @ 100% FI02
High Flow
PM5580: 80/20 Heliox
PM5570: 70/30 Heliox
13 l/min or less at 50 psi (3.4 bar)

Low Flow
PM5480: 80/20 Heliox
PM5470: 70/30 Heliox
3 l/min or less at 50 psi (3.4 bar)

Maximum Combined Flow
(All Outlets)
High Flow
PM5580: 80/20 Heliox
PM5570: 70/30 Heliox
≥ 120 l/min

Low Flow
PM5480: 80/20 Heliox
PM5470: 70/30 Heliox
≥ 30 l/min

Bypass Flow
(Loss of Helium or Oxygen supply)
High Flow
PM5580: 80/20 Heliox
PM5570: 70/30 Heliox
≥ 85 l/min

Low Flow
PM5480: 80/20 Heliox
PM5470: 70/30 Heliox
≥ 45 l/min

Bypass Alarm Activation
High Flow
PM5580: 80/20 Heliox
PM5570: 70/30 Heliox – 50 psi (3.45 bar)
13 – 25 psi (0.9 – 1.7 bar)

High Flow
PM5580: 80/20 Heliox
PM5570: 70/30 Heliox – 60 psi (4.14 bar)
16 – 24 psi (1.1 – 1.5 bar)

Low Flow
PM5480: 80/20 Heliox
PM5470: 70/30 Heliox – 50 psi (3.45 bar)
18 – 22 psi (1.2 – 1.5 bar)

Low Flow
PM5480: 80/20 Heliox
PM5470: 70/30 Heliox – 60 psi (4.14 bar)
16 – 24 psi (1.1 – 1.65 bar)

Alarm Reset
When pressure differential is 6 psi (0.4 bar) or less.
Alarm Sound Level
≥ to 80 db at 1 ft (0.3 m)
Oxygen Concentration Adjustment Range
PM5480 / PM5580
20 – 100%PM5470 / PM5570
30 – 100%
Gas Supply Pressure
30 – 75 psi (2.1 – 5.2 bar) Heliox & Oxygen within 10 psi (0.69 bar) of each other
Mixed Gas Stability
± 1% Oxygen
Connection Types
DISS Type – Heliox & Oxygen Inlets & Outlets and/or NIST Type – Heliox & Oxygen Inlets
Note: All flow – rate values are as measured from an Oxygen flowmeter (uncorrected).
Dimensions (without fittings)
Depth: 4.9 in (12.5 cm)
Width: 2.3 in (5.7 cm)
Height: 4.1 in (10.4 cm)
Weight
Unit Weight: 2.29 lbs (1.04 kg)Shipping Weight: 2.95 lbs (1.34 kg)
Operating Temperature Range
59°F to 104°F (15°C to 40°C)
Transport/Storage Requirements
Temperature Range: -10°F to 140°F (-23°C to 60°C)
Humidity: Max 95% Non-condensing
FI02 Accuracy*: ± 3% of full scale @ 50 psi (3.4 bar)
Pressure Drop
Low Flow: ≤ 2 psi (0.14 bar)at inlet pressures from 30 – 90 psi (2.1 – 6.2 bar) and at 10 l/min flow rate at 60% FI02.
High Flow: ≤ 3 psi (0.14 bar)at inlet pressures from 30 – 90 psi (2.1 – 6.2 bar) and at 10 L/min flow rate at 60% FI02.
Pressure Drop Additional Info
The Helium-Oxygen Blender has been cleaned for Oxygen Service prior to delivery.The Helium-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.

The Oxygen Analyzer should comply with ISO 21647.

Dryness and Composition for inlet gases
Heliox: Medical grade 80/20 or 70/30 required.

Oxygen: Oxygen supply must meet all requirements of USP Medical Grade Oxygen.

Dew Point (ONLY for CE requirements): Both inlets should remain 10°F (5.55°C) or more below the lowest temperature to which the air distribution system equipment is exposed. At a temperature of 25°F (-3.9°C) and a pressure of 90 psi (6.33 kg/cm2) this equates to 2000 mg/m3.

*Accuracy of FI02 will be affected if bleed flow is not engaged at low flows. (At or below 3 l/min for Low Flow and 15 l/min for High Flow). When Heliox tank pressure and oxygen outlet pressures are unbalanced, bleed may need to be engaged at a higher liter flow to maintain accuracy.

Specifications are subject to change without prior notice.

LIMITED WARRANTY AND LIMITATION OF LIABILITY

Precision Medical, Inc. warrants that the HeliO2 Blender, (the Product), will be free of defects in workmanship and/or material for the following period:

Two (2) years from shipment

Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.

ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.

The representatives of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.

THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.

Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.

Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.